Knife-Edge Beam Width Calculator
How wide is a beam that a blade has been scanned across? Enter the blade travel between two power levels, such as 10 % and 90 %, or paste the whole scan, and the calculator returns the 1/e2 radius, the diameter, the FWHM and the beam center, with the fitted error function drawn over the points. Background: knife-edge measurement, beam waist and beam diameter.
Two short experiments. Each one sets the inputs, says where to look, and asks for a prediction before it shows the result.
These checks run in your browser on every load. The clip-width factors are compared with the standard normal quantiles, and the fit with a scan computed for known parameters.
| Check | Expected | Computed | Tolerance |
|---|
The expected values follow P(x) = (P0/2) erfc(√2 (x − x0)/w) for a Gaussian beam, evaluated by hand or from tabulated quantiles. The tolerance is the largest difference from Expected that still passes, relative to Expected or to 1, whichever is larger.
A blade at position blocks one side of a Gaussian beam, and the power that passes is
where is the unblocked power, a background or detector offset, the beam center and the 1/e2 intensity radius. The intensity profile along the scan is a Gaussian with standard deviation , so the blade positions where a fraction and of the power passes are from the center, with the standard normal quantile. The clip width is therefore
and the 15.87 % to 84.13 % width equals exactly. A pasted scan is fitted for all four parameters by Levenberg-Marquardt least squares, which uses every point rather than two interpolated crossings and reports the RMS residual as a check on the Gaussian assumption. The model is one-dimensional: a knife edge integrates over the other axis, so an elliptical beam needs a scan along each axis, and a multimode or clipped beam gives a fitted that is no longer its second-moment radius.
Worked example
A 10–90 % blade travel of 64.08 µm gives , a 1/e2 diameter of 100.0 µm and an intensity FWHM of . The example scan, 21 points 10 µm apart computed for a 50.2 µm beam with a little noise added, fits to with the center at 4.175 µm and an RMS residual of 0.2 % of the step; reading the 10 % and 90 % crossings from the same points by straight-line interpolation gives 50.93 µm, which is why the fit is the better estimate when the points are coarse.
Related: the knife-edge measurement entry describes the setup, the M² measurement procedure uses widths taken at several distances, and the Gaussian beam calculator propagates the radius found here.
References: J. M. Khosrofian and B. A. Garetz, “Measurement of a Gaussian laser beam diameter through the direct inversion of knife-edge data,” Applied Optics 22, 3406–3410 (1983). ISO 11146-1:2021, Lasers and laser-related equipment: test methods for laser beam widths, divergence angles and beam propagation ratios. A. E. Siegman, Lasers, University Science Books (1986), chapter 17.